4.16 Service menu
4-48
where:
Work time
parameter allows to see the whole operating time of the device.
Time after maintenance
parameter allows to control time after the previous maintenance
of the device for the next timely maintenance. For this purpose the expert who was
carrying out maintenance, should choose the “Zero time after maintenance” parameter
and then the time counter returns to starting (zero) position.
Clear alarm log
parameter allows to clear information from the memory. For the
activation of this action enter the system password. For exit without emptying the alarm
log press the encoder for 4 times.
Transfer service log
parameter allows to make record of maintenance log to the portable
data medium connected to USB-port. The maintenance log is saved in "servicelog.csv"
file, in the csv format with the symbol divider ";", in the CP866 coding.
Technological mode
parameter is intended for the transfer of the device to the mode of
the gas leakage test in the breathing circuit being under pressure during inspiratory time,
and also test of pressure drop of the device in the passive exhalation line (the specified
tests are carried out by the manufacturer and in the service centers, for these tests it is
necessary to enter the system password).
Technological child mode
parameter is intended for the test of device by the
manufacturer and in the service centers.
Service information
parameter is intended for technical staff (to perform the operation it
is necessary to enter the system password).
•
Calibration
parameter allows to enter submenu including the following parameters:
o
CIVL
o
CIND
o
Breathing circuit calibration
o
CAPNO (mainstream capnometer)
o
SpO
2
module (option).
CIVL
parameter is needed for entering to the menu of calibration selection of the
ventilation controller.
Table 4.10. CIVL calibration parameters
:
Parameter
Function
Description
FiO
2
sensor calibration
Call of procedure of the oxygen sensor calibration
it.4.16.2
FiO
2
sensor replacement
Call of procedure of the oxygen sensor replacement
it.4.16.2
Exp. flow sens. calibr.
Call of procedure of expiratory flow sensor
it.4.16.3
Summary of Contents for MV200
Page 2: ......
Page 8: ......
Page 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Page 124: ...4 21 Open valve function 4 76 For your notes ...
Page 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Page 160: ...7 Troubleshooting 7 2 For your notes ...
Page 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Page 166: ...13 commission date mark 12 2 For your notes ...
Page 168: ...14 Maintenance and repair data 14 2 For your notes ...
Page 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Page 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Page 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Page 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...