![Treaton MV200 User Manual Download Page 16](http://html1.mh-extra.com/html/treaton/mv200/mv200_user-manual_1154868016.webp)
1.4 Information about resource of the main functional units of the device
1-8
1.4 Information about resource of the main functional units of the device
The device has the built-in flow generator providing its independence from sources of com-
pressed air. The design and completing units provide guaranteed operability of the built-in flow
generator during at least 40 000 operating hours, or during the whole period of operation of 10
years (the smallest parameter is chosen). Operability of the built-in flow generator is guaranteed
only at following operating and service conditions stated in the present document.
For the monitoring of oxygen concentration in the inhaled air the device has the built-in electro-
chemical oxygen sensor with the operation resource at least 10 thousand hours.
For the service control the device has the special operating time counter. After 6 000 operating
hours the message appears at the screen of the device “Carry out maintenance”. The engineer
who has carried out maintenance should enter the command in the service menu “Reset time
after maintenance” then the time counter will return to the starting position.
1.5 Revision history
Every version of User manual has its number and date of issue. Version number is changed
with the significant technical changes of the device. When issuing new version, date of version
is also changed. Insignificant amendments and additions that don’t influence on the modes,
functions and parameters of ventilator, don’t cause the change of version date.
Summary of Contents for MV200
Page 2: ......
Page 8: ......
Page 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Page 124: ...4 21 Open valve function 4 76 For your notes ...
Page 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Page 160: ...7 Troubleshooting 7 2 For your notes ...
Page 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Page 166: ...13 commission date mark 12 2 For your notes ...
Page 168: ...14 Maintenance and repair data 14 2 For your notes ...
Page 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Page 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Page 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Page 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...