2.1 Basic parameters and characteristics
2-2
No
Parameter
Value (description)
2.2
Continuous mandatory ventilation with the controlled
pressure of inspiration (synchronized ventilation with
pressure control)
CMV/PCV
2.3
Synchronized intermittent mandatory ventilation with
flow trigger or pressure trigger with volume control
with pressure support of spontaneous breaths (PS)
SIMV/VC
2.4
Synchronized intermittent mandatory ventilation with
flow trigger or pressure trigger with pressure control
with pressure support of spontaneous breaths (PS)
SIMV/PC
2.5
Synchronized intermittent mandatory ventilation with
flow trigger or pressure trigger with double control with
pressure support of spontaneous breaths (PS)
SIMV/DC
2.6
Spontaneous breathing with continuous positive air-
way pressure with pressure support of spontaneous
breaths (PS)
CPAP+PS
2.7
Spontaneous breathing with two levels of continuous
positive airway pressure with pressure support of
spontaneous breaths (PS)
BiSTEP
2.8 Airway pressure release ventilation
APRV
2.9
Pressure controlled ventilation with guaranteed respi-
ratory volume
PCV-VG
2.10 Non-invasive ventilation of lungs
NIV
2.11 Intellectual support ventilation (option)
iSV
2.12
Automatic backup ventilation mode in cases of pres-
sure and volume apnea.
APNEA
3
Ventilation parameters
3.1
Tidal volume, Vt
Measuring accuracy :
In range 10-100 ml
In range 100-3000 ml
10 - 3000 ml
±(2+0.05Vt) ml abs
±8 % rel
3.2 Minute volume, MV
0 - 60 lpm
3.3
Rate of breathing / respiratory rate (controlled respira-
tory cycles), RB:
1 - 120 1/min
3.4
Frequency of mandatory breaths at synchronized in-
termittent assisted ventilation, RB
1 - 60 1/min
3.5
Waveform of the gas flow in the ventilation modes with
volume control, FormFlow
Rectangular, descending
3.6 Periodic enhanced inspiration
Available
3.7 I:E ratio
1:99 - 60:1
3.8 Plateau time, Tplat
0 - 5 s
0 - 70 % from inspiratory time
3.9 Support pressure of spontaneous breath, PS
0 - 80 cmH
2
O (mbar)
3.10 Positive end-expiratory pressure, PEEP
0 - 50 cmH
2
O (mbar)
3.11 Inspiratory pressure, Pi
0 - 100 cmH
2
O (mbar)
3.12 Inspiratory time, Tinsp
0.2 - 10 s
3.13 Trigger window, TrigWnd
0 - 100 %
0.5 - 4 s
3.14 Flow trigger sensitivity, Ftrig
0.5 - 20 lpm
3.15 Pressure trigger sensitivity, Ptrig
0.5 -20 cmH
2
O (mbar)
Summary of Contents for MV200
Page 2: ......
Page 8: ......
Page 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Page 124: ...4 21 Open valve function 4 76 For your notes ...
Page 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Page 160: ...7 Troubleshooting 7 2 For your notes ...
Page 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Page 166: ...13 commission date mark 12 2 For your notes ...
Page 168: ...14 Maintenance and repair data 14 2 For your notes ...
Page 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Page 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Page 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Page 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...