7. Troubleshooting
7-1
7
TROUBLESHOOTING
Name of the fault, external dem-
onstrations and additional signs
Probable cause
Remedies and recommenda-
tions for further actions
The device does not work from the
mains power supply,
"
Power
"
LED
is not illuminated. When operated
from the accumulator the alarm
message about the absence of
mains voltage is displayed.
No mains voltage or voltage is
low
Check the mains voltage in outlet
Fuses have burned out
Check or replace the mains fuses
The device does not operate from
the accumulator, the alarm mes-
sage about accumulator discharge
is displayed.
Accumulator is discharged
Charge the accumulator (see it.
4.2)
"Battery" indicator is red
Accumulator or charger is faulty Call for Triton Electronic Systems
Ltd. service or your local repre-
sentative
Alarm message about disconnec-
tion is displayed
Disconnection of the circuit has
happened
Check the tightness of the
breathing circuit connections; if
necessary replace the breathing
circuit
Alarm message about occlusion is
displayed
Breathing circuit, inspiration or
expiration filter is occluded.
Check breathing circuit for occlu-
sion or compression, check the
inspiration / expiration filters, if
necessary clean
Faulty expiration valve
Call for Triton Electronic Systems
Ltd. service or your local repre-
sentative
Alarm message about high oxygen
pressure is displayed
The device is connected to the
compressed
oxygen
source
without the filter-regulator (or fil-
ter-regulator is defective), and
safety valve was triggered (0.25
- 0.35 MPa)
Connect or replace the filter-
regulator
Alarm message about low oxygen
pressure is displayed
Input pressure of oxygen line is
lower than 0.2 MPa
Ensure the normal operating
pressure on the device input.
The device has switched into the
emergency mode. Alarm message
about ventilator replacement and
error codes are displayed
Technical failure of one or more
modules of the device.
Call for Triton Electronic Systems
Ltd. service or your local repre-
sentative
"Fan malfunction" alarm message
is displayed
Fan malfunction (blades don’t
rotate)
Call for Triton Electronic Systems
Ltd. service or your local repre-
sentative
If the problem can not be solved by the proposed measures, contact your ser-
vice representative, or customer service of the manufacturer Triton Electronic
Systems Ltd.
Summary of Contents for MV200
Page 2: ......
Page 8: ......
Page 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Page 124: ...4 21 Open valve function 4 76 For your notes ...
Page 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Page 160: ...7 Troubleshooting 7 2 For your notes ...
Page 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Page 166: ...13 commission date mark 12 2 For your notes ...
Page 168: ...14 Maintenance and repair data 14 2 For your notes ...
Page 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Page 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Page 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Page 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...