4.19 Features of using virus-bacterial and heat and moisture exchange filters
4-71
At opening of the extended respiratory monitoring windows the alarm list
becomes unavailable. To access the alarm list, go to the RESP 1 window.
4.19 Features of using virus-bacterial and heat and moisture exchange
filters
When
antibacterial filter
is in the expiration line the undesirable increase in breathing resis-
tance can occur, especially if the filter is used together with the nebulizer and humidifier. This
can lead to the hard work of the patient's breathing and increased pressure in the respiratory
system (autoPEEP phenomenon). Unwanted pressure increase in the respiratory system of the
patient can be identified by the fact that the flow of exhaled gas by the end of exhalation does
not fall to zero.
Check the virus-bacterial filter and replace it, if it is suspected that is causes autoPEEP increas-
ing.
Figure 4.12 – The effect of extending of the expiraiton flow due to the increased filter resistance
The heat and moisture exchange filter
in the breathing circuit can significantly increase resis-
tance to the inspiration and expiration.
Increased resistance causes the increased patient’s work during spontaneous breathing or in-
itialization trigger effort at trigger support ventilation. Under unfavorable circumstances, increas-
ing of the expiration resistance can lead to an undesirable increasing of the pressure in the res-
piratory system of the patient (autoPEEP phenomenon).
The resistance connected with the heat and moisture exchange filter, the device cannot control
directly. Therefore, when using the heat and moisture exchange filter user should:
•
frequently monitor the patient's condition;
•
carefully follow to the guidance on the proper use of heat and moisture exchange filter.
It is prohibited to use heat and moisture exchange filter with nebulizer or
with the respiratory gas humidifier!
Summary of Contents for MV200
Page 2: ......
Page 8: ......
Page 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Page 124: ...4 21 Open valve function 4 76 For your notes ...
Page 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Page 160: ...7 Troubleshooting 7 2 For your notes ...
Page 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Page 166: ...13 commission date mark 12 2 For your notes ...
Page 168: ...14 Maintenance and repair data 14 2 For your notes ...
Page 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Page 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Page 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Page 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...