5.5 Metabolism measurement module
5-17
No Alarm message
Alarm conditions
Reaction of ventilator,
operator’s actions
5. Drying
Drying of the line
The message disappeared automatically af-
ter calibration finishing.
6. User turn off
Turned off by user
The message disappeared automatically af-
ter turning on the metabolism module.
7. Off. Occlusion
Turned off due to occlusion
The message disappeared automatically af-
ter removing the occlusion and turning on
the metabolism module.
8. Emitter malfun
Emitter malfununction
The ventilator continue operation without
metabolism module.
Contact service department.
9. Receiver malf.
Receiver malfununction.
10. Compress malf.
Compress malfununction.
11. Press.sens err
Pressure sensor error
12. Off. Overheat
Off. Overheat
13. Poured chamber
Emitter malfununction
14. Not used
Metabolism module is not
used
Metabolism module is not used. The mes-
sage disappeared automatically after start-
ing operation.
15. Blowing failed
Blowing failed
The message disappeared automatically af-
ter finishing of the blowing.
Correct measuring of metabolic parameters is provided at the FiO
2
values
not exceeding 60 %. This is caused by the features of the current method
of measurement.
For the more accurate measurement of metabolic parameters it is recom-
mended before the start of patient’s ventilation to calibrate the O
2
sensor
of metabolism measurement module through [Menu]
→
→
→
→
[Service Menu]
→
→
→
→
[Calibration]
→
→
→
→
[CAPNO] in accordance with it. 4.16.2.
•
Calculation of metabolic parameters is blocked in the absence of fixing of respiration, during
zero calibration and occlusion of sampling line.
•
When operating with metabolism measurement module remember that explosion of liquid
through sampling line in its measuring path can affect normal operation, so the device has
number of measures for protection against moisture.
•
The main barrier protecting the measuring path from the liquids is water trap for the separa-
tion and condensation of the moisture in the sample and for its accumulation in the special
removable water trap reservoir, which periodically has to be emptied. It takes about 40 hours
for water trap to become full under condition of operating temperature 23°C, patient respira-
tion gas 37° and 100% RH.
Handle the contents of water trap as you would handle any bodily fluid.
Failure to follow this instruction causes infection.
•
During long-term operation of metabolism measurement module under high humidity of the
sample gas sampling line can accumulate a significant amount of condensate, which leads to
the formation of liquid droplets. Large number of droplets can cause occlusion of the measur-
ing path and metabolism measurement module malfunction. In the case of frequent messag-
es about tract occlusion it is usually sufficient to replace the sampling line for a new dry one.
•
After every finishing of operation with metabolism measurement module it is recommended
to disconnect the sampling line from it and run for 5 ... 10 minutes to dry its internal pneumat-
ic system, and to extend the service life of the oxygen sensor.
Summary of Contents for MV200
Page 2: ......
Page 8: ......
Page 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Page 124: ...4 21 Open valve function 4 76 For your notes ...
Page 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Page 160: ...7 Troubleshooting 7 2 For your notes ...
Page 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Page 166: ...13 commission date mark 12 2 For your notes ...
Page 168: ...14 Maintenance and repair data 14 2 For your notes ...
Page 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Page 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Page 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Page 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...