6.7 Check and calibration of the FiO2 oxygen sensors and the metabolism measurement module
6-4
After each disassembly / assembly of the expiration valve, it is recommended to calibrate flow
sensor at the expiration (it. 4.16.3).
6.7 Check and calibration of the FiO
2
oxygen sensors and the metabol-
ism measurement module
The device contains two oxygen sensors - the FiO2 oxygen sensor and the oxygen sensor of
the metabolism measurement module.
To verify the accuracy of measuring oxygen sensors make the following:
•
Connect the device to the oxygen source.
•
Assemble breathing circuit. Connect patient tee to the respiratory bag through intubation
tube.
•
Connect water trap of ventilator with patient tee port by the sampling line.
•
Switch on the device in any mandatory ventilation mode, set the tidal volume V
T
of 400 ml.
•
Monitor the FiO
2
oxygen sensor readings in one of the measurement parameter blocks (by
selecting the value "FiO
2
"). The readings of the metabolism measurement module’s oxygen
sensor can be monitored in the respiratory monitoring window "Metabolism" (can be selected
using the graphic button "Monitoring").
•
By setting different FiO
2
values in the range 21 - 100%, check that the absolute error of the
measured values of FiO
2
does not exceed ± 3%.
If the measurement error exceeds ± 3%, disconnect the device from the oxygen source and ca-
librate the oxygen sensors (it. 4.16.2).
If calibration fails, replace oxygen sensors.
Replacement of the oxygen sensor of metabolism measurement module requires disassembly
of the device and can only be performed by qualified personnel authorized by Triton Electronic
Systems Ltd.
Replacement of the FiO
2
oxygen sensor can be performed independently. For this purpose un-
screw the two screws of the oxygen sensor cover. Carefully take the inspiratory connector as-
sembly, so as not to break off the wires of the connector. Disconnect the sensor power cable.
By counterclockwise rotating unscrew the faulty sensor (Figure 6.4). Install a new sensor (in
some cases, a spare sensor can be included in the delivery kit). Assemble the inspiratory con-
nector in the reverse order. Calibrate the oxygen sensor (item 4.16.2).
Figure 6.4 – FiO
2
sensor replacement
Disposal of electrochemical oxygen sensors shall be performed similarly to batte-
ries and accumulators by collection in specialized containers for further reprocess-
ing according national rules.
Summary of Contents for MV200
Page 2: ......
Page 8: ......
Page 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Page 124: ...4 21 Open valve function 4 76 For your notes ...
Page 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Page 160: ...7 Troubleshooting 7 2 For your notes ...
Page 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Page 166: ...13 commission date mark 12 2 For your notes ...
Page 168: ...14 Maintenance and repair data 14 2 For your notes ...
Page 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Page 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Page 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Page 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...