5.7 “Auxiliary external pressure” function
5-23
a)
b)
Figure 5.21 – Connection of the endotracheal tube (a) and the tee of the breathing circuit (b)
Assembly of the circuit of external auxiliary pressure measurement in esophagus or
stomach
•
Follow the procedure operation instructions supplied with the catheters.
•
Connect a hydrophobic (virus-bacterial) filter with the Luer connector and monitoring line to
the Paux pressure port (Figure 5.22).
Figure 5.22 - Connection of the esophageal balloon catheter or multifunction nasogastric cathe-
ter to the ventilator
Indication of transpulmonary pressure
To display the calculated transpulmonary pressure select the Ptp value for one of the mea-
surement parameter blocks through [Menu]
→
[Display settings]
→
[Choose measured
par.blocks]:
Summary of Contents for MV200
Page 2: ......
Page 8: ......
Page 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Page 124: ...4 21 Open valve function 4 76 For your notes ...
Page 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Page 160: ...7 Troubleshooting 7 2 For your notes ...
Page 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Page 166: ...13 commission date mark 12 2 For your notes ...
Page 168: ...14 Maintenance and repair data 14 2 For your notes ...
Page 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Page 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Page 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Page 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...