ー 5 ー
WARNING
Make sure that holes inside the chamber
are not blocked by articles to be sterilized.
If the vent hole is blocked by such items
as sterilization bags, the chamber pressure
cannot be controlled, thereby causing
serious accident such as explosion of the
chamber. Make sure that articles to be
sterilized are placed in a container, such
as a basket provided with the FLS-1000 to
prevent the vent hole and its surrounding
area from being blocked.
Stop operation immediately when the
pressure gauge pointer is over the red
line.
Otherwise, parts can be damaged and
broken parts may scatter, resulting in burns,
injury or serious accident. In this case, stop
operation immediately and contact your
dealer or our local representative.
Use special caution when handling the
articles to be sterilized that may be very
hot after the operation.
Handling hot articles can cause burns.
Use special caution when handling the
hot liquid that may suddenly boil over.
The sterilized liquid inside the chamber is
cooled down slowly than the chamber and
may boil over suddenly on any shocks. This
bumping may cause burns or unexpected
accident.
Do not leave any contamination inside
the chamber or in the sterilizing water.
Failure to do so may cause corrosion or
damage on the chamber. It may also cause
malfunction of the water level sensor,
which may hamper the empty heating
prevention function, resulting in fire.
Be sure to unplug the FLS-1000 before
performing maintenance procedures.
Failure to do so may cause electric shock.
For models with no power plug provided,
the power circuit breaker to which the
power plug is connected should be turned
off.
Do not operate the FLS-1000 if any
corrosion, damage, deformation or loose
connections are found on its chamber,
chamber lid, hinges, hinge shafts,
retaining rings, lid hooks, or lid hook
catches.
Otherwise, explosion may occur while the
pressure rises, resulting in death, injury
or serious accident. If any abnormality is
found, please contact your dealer or our
local representative.
Do not operate the FLS-1000 if any crack,
deformation or damage is found on its
chamber lid gasket.
Otherwise, steam leak may occur, resulting
in burns. If any abnormality is found,
please contact your dealer or our local
representative.
After replacing the chamber lid gasket,
be sure to make a test run.
If the chamber lid gasket is not installed
properly, steam may leak and cause burns.
Unplug and inspect the power plug
regularly.
Failure to do so may cause fire or burnout.
Do not operate the FLS-1000 if the
chamber lid handle is deformed or
damaged.
Failure to do so may cause burns or injury.
If any abnormality is found, please contact
your dealer or our local representative.
Do not attempt to open the drain cock
unless the pressure gauge reads 0MPa.
If the drain cock is opened when pressure
still remains in the chamber, hot water may
spurt out, causing death, injury or serious
accident.