ー 47 ー
Press the FUNCTION key while the IN PROGRESS
indicator is off. The time display flashes “F00”.
Press the time setting up and down arrow keys
to show the course number to be changed. The
temperature display shows the currently set value
of the lid interlock release temperature.
※
Select F09 and F11 for liquid sterilization course and
sterilization course, respectively.
Press the START key. The temperature display starts to flash.
Press the temperature setting up and down arrow keys to show the desired value for the
temperature of the lid interlock release.
Press and hold down the START key.
The buzzer will sound in approximately 4 seconds to indicate that the setting is finalized and
the display will return to normal display.
※
The factory default values for both F09 and F11 are set at 97℃.
※
The upper limit and specified value may vary depending on the altitude level. (Refer to “3-3.
Setting the Altitude Level”.)
※
The lid interlock release temperature can also be monitored by the external temperature sensor for
articles to be sterilized.
(
Refer to “7-1. Temperature Sensor for Articles to be Sterilized”.
)
①
②
③
④
⑤
6-9.
Setting the Temperature to Release the Lid Interlock
The FLS-1000 is equipped with the lid interlock which activates and prevents the lid from
opening when the lid is forced open while the chamber temperature is still high. The lid
interlock release temperature can be changed depending on the operation. It can be separately
set for the sterilization course and liquid sterilization course.
STOP
MEMORY
STERILIZE
LIQUID
START
FUNCTION
CHECK
TIMER
EXT.
SENSOR
COOLING
FAN
EXHAUST SPEED
FAST
TIME
TEMPERATURE
TIMER
READY
STERILIZE
LID
LID
WATER
DRAIN
COVER
CHANGE WATER
°C
SLOW
STERILIZE
IN PROGRESS
STOP
MEMORY
STERILIZE
LIQUID
START
FUNCTION
CHECK
TIMER
EXT.
SENSOR
COOLING
FAN
EXHAUST SPEED
FAST
TIME
TEMPERATURE
TIMER
READY
STERILIZE
LID
LID
WATER
DRAIN
COVER
CHANGE WATER
°C
SLOW
STERILIZE
IN PROGRESS
Summary of Contents for FLS-1000
Page 9: ...ー 8 ー MEMO ...
Page 10: ...ー 9 ー 2 2 Parts Identifications FLS 1000 Main Unit ...
Page 51: ...ー 50 ー MEMO ...