ー 33 ー
Press the STERILIZE key.
The STERILIZE indicator lights up to indicate that
the sterilization course is selected, while all process
indicators on the work monitor light up green.
Press the temperature setting up and down arrow
keys to set the temperature for the sterilization
process.
Press the time setting up and down arrow keys to
set the time for the sterilization process.
Press the exhaust speed level setting up and down arrow keys to set the exhaust level for the
temperature and pressure falling process.
Press the cooling fan setting key to select the number of fans which operate during the
temperature and pressure falling and temperature falling processes.
<
To set the sterilization operating parameters
>
①
②
③
④
⑤
5-7. Setting Operating Parameters
※
To change the time display format, refer to “6-7. Setting the Time Display Format”.
※
The upper limit of the temperature setting range varies depending on the altitude level selected
at the time of installation of the autoclave.
Altitude level code
Altitude level
Sterilization process
Temp. setting range
Time setting range
0
Below 800m
105
〜
123℃
0 : 01
〜
9 : 59
(switchable:
〜
99 hours)
1
Above 800m
〜
below 1500m
105
〜
122℃
2
Above 1500m
〜
below 2000m
105
〜
121℃
Table. Altitude level code and the corresponding altitude level, temperature and time setting ranges in
sterilization process
STOP
MEMORY
STERILIZE
LIQUID
START
FUNCTION
CHECK
TIMER
EXT.
SENSOR
COOLING
FAN
EXHAUST SPEED
FAST
TIME
TEMPERATURE
TIMER
READY
STERILIZE
LID
LID
WATER
DRAIN
COVER
CHANGE WATER
°C
SLOW
STERILIZE
IN PROGRESS