ー 39 ー
5-11. Draining the Sterilizing Water from the Chamber
CAUTION
Allow sterilizing water to cool down sufficiently before draining water
from the chamber.
●
Water in the chamber or the exhaust tank may be very hot after the operation,
causing burns.
①
②
③
④
⑤
⑥
⑦
⑧
⑨
LO W E R
U P P E R
Open the chamber lid.
Open the drain access cover.
Pull out the drain hose.
Place the drain container underneath the drain hose.
Open the drain cock by rotating it slowly and drain the
sterilizing water from the drain port. If the water drains very
slow or does not drain, the port may be clogged. Clean the drain
port with tools such as a wire to remove the clog.
After completely draining water from the chamber, close the
drain cock.
Place the drain hose back to its original position.
Close the drain access cover.
Close the chamber lid.
Open
Close
WARNING
Do not attempt to open the drain cock unless the pressure gauge reads
0MPa.
●
If the drain cock is opened when pressure still remains in the chamber, hot
water may spurt out, causing death, injury or serious accident.
Summary of Contents for FLS-1000
Page 9: ...ー 8 ー MEMO ...
Page 10: ...ー 9 ー 2 2 Parts Identifications FLS 1000 Main Unit ...
Page 51: ...ー 50 ー MEMO ...