ー 19 ー
To move the FLS-1000 hold the top of the unit on both sides of the chamber lid handle with
both hands and carefully push it.
※
Four casters are mounted in the four corners of the bottom of the FLS-1000. A caster in the corner
where the chamber lid handle is attached on top is a fixed one, and casters on each of the other
three corners are swivel ones.
※
Never attempt to lift the FLS-1000 while it is being moved. This model is not designed to be lifted
up.
①
<
To move and install the FLS-1000
>
To install the FLS-1000 select a location in a room that is well ventilated and free of direct
sunlight.
Place the FLS-1000 onto a solid flat surface such as a concrete floor.
※
Weight of the FLS-1000 : 108kg
Avoid installing the FLS-1000 in a slippery place or a place which may be exposed to
vibration.
Considering the steam coming from steam exhaust vent and surrounding area, install
FLS-1000 providing a clearance of at least 25cm at the rear and on each side of the unit.
When operating the FLS-1000, the following environmental requirements should be met:
○
○
○
○
○
<
Installation Location
>
: 10 to 35°C
: 750 to 1060hPa
: less or equal to 75
%
: below 2,000m
: 2 degrees
: Class II
: 2
Ambient temperature range
Atmospheric pressure range
Relative humidity
Altitude
Maximum gradient
Overvoltage category
Pollution degree