ー 38 ー
5-10. Removing Articles to be Sterilized
WARNING
Never attempt to open the lid until the chamber pressure returns to
0MPa.
●
If the lid is opened while pressure remains in the chamber, articles to be
sterilized or steam may gush out, thereby resulting in death or serious injury.
Be extremely careful of steam coming out of the chamber when
opening the chamber lid.
●
Steam coming out of the chamber is very hot and can cause burns or injury.
Use special caution when handling the hot liquid that may
suddenly boil over.
●
The sterilized liquid inside the chamber is cooled down slowly than the
chamber and may boil over suddenly on any shocks. This bumping may
cause burns or unexpected accident.
Do not leave any contamination inside the chamber or in the
sterilizing water.
●
Failure to do so may cause corrosion or damage on the chamber. It may also
cause malfunction of the water level sensor, which may hamper the empty
heating prevention function, resulting in fire.
Check that the gauge for pressure in the chamber reads “0MPa”.
Check that the lid interlocked indicator is off.
Open the chamber lid.
Remove articles to be sterilized from the chamber.
※
When removing articles to be sterilized, make sure that they do not touch the chamber lid gasket
or the temperature sensor. It may cause damage.
Clean if the chamber interior or sterilization water is contaminated.
※
Refer to “8. ROUTINE MAINTENANCE”.
①
②
③
④
⑤
CAUTION
Be careful of the chamber or its surrounding area after the operation
because it is very hot.
●
Failure to do so may cause burns.