ー 30 ー
5-5. Checking the Sterilizing Water in Chamber
Open the chamber lid and check that the sterilizing
water in the chamber is not contaminated. Drain
water from the chamber if the sterilizing water is
contaminated.
※
For drainage of water, please refer to “5-11. Draining
the Sterilizing Water from the Chamber”.
Check that the water level in the chamber is suffi cient
enough to cover the water level indicator. Add water
to the chamber, if necessary, until the indicator is
completely covered with water.
※
The water level in the chamber should be maintained
only barely higher than the surface of the water level
indicator to prevent articles to be sterilized from
getting wet.
※
To sterilize the large amount of liquid in a bucket, fi ll
the chamber with water until the bottom of the bucket
is covered to approximately 5 cm.
Check the low sterilizing water level indicator turns
off .
<
To check the quality and amount of sterilizing water
>
①
②
③
The FLS-1000 generates steam by heating up the sterilizing water in the chamber, which has
a heating element in its bottom. Check the quality and level of the sterilizing water inside the
chamber to perform adequate sterilization.
Use tap water as sterilizing water. When purifi ed water or cold water is used, the water level
sensor may not correctly detect the level of the sterilizing water in the chamber.
If tap water has inferior quality, purifi ed water may be added to fi ll the chamber. In such a
case, fi ll the chamber with purifi ed water fi rst until the water level indicator is covered in
water and then, fi ll it with tap water until the low sterilizing water level indicator turns off .
Quality of tap water may vary widely, depending on location. Therefore, it is recommended that
approximately 300mL of tap water per 3L of purifi ed water be used as a guide.
•Up to 5cm above the bottom of the bucket
approx. 6.7L
•Up to the chamber bottom plate
approx. 6.0L
•Up to the water level indicator
approx. 4.4L
<
Approximate indication of the amount of water in the chamber
>
5㎝
Bucket
Water Level Indicator