ー 58 ー
Error code
Possible cause
Remedy
A01
The start key is pressed before draining and
pouring water into the chamber.
※
The default value is 40 cycles.
Refer to “6-10. Sterilizing water change Alert
Function”.
A02
The start key is pressed when the water
level in the chamber is low.
※
Low sterilizing water level is detected.
Fill the chamber with water.
A03
The start key is pressed when the lid open/
close system failed to detect the lid status.
※
1
Open then close the lid.
A04
The start key is pressed when the lid remains
open.
Close the lid.
A05
The start key is pressed when the drain
access cover remains open.
The drain access cover is opened during the
operation.
Close the drain access cover.
A06
The start key is pressed when the exhaust
tank is full.
Drain the exhaust tank.
A07
The start key is pressed when the
temperature inside the chamber is below
0℃.
Wait until the chamber temperature rises to the
operating temperature.
A09
Power-off is detected after a sterilization
process has started or before it has completed.
Refer to “6-5. Power Failure Alert Function”.
A10
The memory approaches the end of its
lifetime. (E26)
※
An alarm notification appears when
80
%
, 90
%
, 95
%
, 98
%
, 99
%
, and 99.5
%
of the lifetime is reached.
Press the check key once to clear the error, and
enable the system.
A20
The start key is pressed while printing.
Press the check key once.
A21
Printer trouble.
※
Printer error includes: cover open, paper
out, printer head abnormal temperature,
and disconnection.
Refer to “7. Optional Functions”.
A30
The calendar battery has reached the end of
its lifetime.
Press the check key once to clear the error, and
enable the system.
Contact your dealer or our local representative.
CHECK
Timing of the annual voluntary inspection
approaches.
※
Reminder activates 30 days before the
due date, and on and after the day of the
due date.
Press the check key once to clear the error,
and enable the system. Refer to “6-6. Annual
Inspection Reminder Function”.
<
Operation Error Code Table
>
※ 1
The FLS-1000 checks the operation of the lid open/close detection system to ensure its safe operation, therefore, open and
close the lid at least once before each operation or timer operation.