7. OPTIONAL FUNCTIONS
The following optional accessories and functions are available for the FLS-1000.
Data output function
The data output function allows producing up to 3 outputs of process data in analog form,
including chamber temperature, temperature of articles to be sterilized, chamber pressure.
The data can be stored to an optional recorder when it is connected.
■
Printer
Data of the chamber temperature, chamber pressure and article temperature can be printed
out on the optional printer.
■
Brackets for anti-tip straps
Metal brackets for anti-tip straps are provided to secure the autoclave with a commercially
available overturn prevention strap or wire.
■
The external temperature sensor for articles to be sterilized
Sterilization process is monitored using the temperature sensor for articles to be sterilized
by installing it directly into the article in the chamber so that the autoclave begins the
sterilization time countdown only when the temperature for articles to be sterilized has
reached a set sterilization temperature and, therefore, provides adequate heating.
※
The temperature is controlled to within 0℃ to
+
2℃ of the set temperature.
※
The lid interlock release and the cycle complete temperatures can also be measured and monitored
by the external temperature sensor for articles to be sterilized.
■
For the operation of the optional accessories and functions, please refer to the supplement to
the Operator’s Manual; “FLS-1000 Autoclave Optional Functions”.