![Tomy FLS-1000 Operator'S Manual Download Page 54](http://html1.mh-extra.com/html/tomy/fls-1000/fls-1000_operators-manual_1126194054.webp)
ー 53 ー
8-2. Weekly Maintenance
WARNING
Do not operate the FLS-1000 if any corrosion, damage, deformation or
loose connections are found on its chamber, chamber lid, hinges, hinge
shafts, retaining rings, lid hooks, or lid hook catches.
●
Otherwise, explosion may occur while the pressure rises, resulting in death,
injury or serious accident. If any abnormality is found, please contact your
dealer or our local representative.
Do not operate the FLS-1000 if any crack, deformation or damage is
found on its chamber lid gasket.
●
Otherwise, steam leak may occur, resulting in burns. If any abnormality is
found, please contact your dealer or our local representative.
Do not operate the FLS-1000 if the chamber lid handle is deformed or
damaged.
●
Failure to do so may cause burns or injury. If any abnormality is found, please
contact your dealer or our local representative.
1. Chamber interior and water level sensor
Remove the chamber bottom plate carefully.
Wipe the chamber interior and the water level sensor using a mild detergent and a cloth or
other suitable materials and wash away any detergent residue using running water and drain
the water.
※
Be careful not to distort sensors in the chamber while cleaning.
<
To clean each part
>
①
②
To ensure safe operation of the FLS-1000, perform cleaning and inspection weekly according to
the following procedures:
2. Chamber lid gasket
Wipe the surface of the chamber lid gasket using a soft clean cloth dipped in clear water and
squeezed almost dry without applying force.
※
Be careful not to damage the chamber lid gasket while cleaning.
Wipe the chamber opening using a soft clean cloth or other suitable materials.
①
②
3. Outer covering
Wipe the outer covering of the unit with a soft cloth moistened with a mild detergent.
Wipe away any detergent residue using a cloth dipped in clear water and squeezed almost dry.
①
②
Unplug and inspect the power plug regularly.
●
Failure to do so may cause fire or burnout.