ー 4 ー
WARNING
Make sure that the FLS-1000 is properly
grounded.
Failure to do so may cause explosion,
electric shock or malfunction.
Never ground the FLS-1000 through a
water pipe or a gas pipe.
Failure to do so may cause explosion,
electric shock or malfunction.
Do not touch the power switch with a
wet hand.
Failure to do so may cause electric shock.
Never attempt to open the lid until the
chamber pressure returns to 0MPa.
If the lid is opened while pressure remains
in the chamber, articles to be sterilized or
steam may gush out, thereby resulting in
death or serious injury.
Do not bend the exhaust hose.
If exhaust venting is not carried out
smoothly, the chamber pressure may
become abnormally high, thereby causing
damage to parts. Broken parts may scatter
and cause burns, injury or serious accident.
When the pressure gauge is not working
properly, stop operation.
Otherwise, abnormality inside the
chamber, if any, cannot be found, thereby
causing serious accident. If pressure gauge
malfunction is suspected, please contact
your dealer or our local representative.
Be extremely careful of steam coming
out of the chamber when opening the
chamber lid.
Steam coming out of the chamber is very
hot and can cause burns or injury.
Do not use the FLS-1000 to sterilize
sealed articles.
Sealed articles may explode or burst when
they are removed from the chamber,
thereby causing burns, injury or serious
accident. The chamber pressure may
become abnormally high, thereby causing
damage to parts. Broken parts may scatter
and cause burns, injury or serious accident.
Do not sterilize any cracked or damaged
glassware.
Damaged glassware may explode or burst
when it is removed from the chamber,
thereby causing burns, injury or serious
accident.
Do not use the FLS-1000 to sterilize
flammable or explosive articles.
Failure to do so may cause fire or explosion.
Do not use extension cords or power
strips.
Failure to do so may cause accidental fire,
short circuit or malfunction.
Avoid leaving any foreign materials
between the chamber lid gasket and the
chamber.
Failure to do so may cause steam leak,
thereby resulting in burns.
Do not use the FLS-1000 to sterilize
corrosive articles.
Otherwise, the chamber may corrode and
explode or burst while pressure rises,
thereby causing death, injury or serious
accident.
Connect the FLS-1000 to a power outlet
of the voltage specified as correct for
FLS-1000. Do not connect it to a shared
outlet.
Failure to do so may cause accidental fire,
short circuit or malfunction.