3-3. Setting the Altitude Level
Press the FUNCTION key while the READY
indicator is on. The time display flashes
“F00”.
Press the time setting up and down arrow
keys to show “F20” in the time display.
The temperature display shows the altitude
level code.
Press the START key.
The temperature display starts to flash.
Press the temperature setting up and down
arrow keys to display the altitude level code
that corresponds to the location where the
FLS-1000 is installed.
<
To set the altitude level
>
①
②
③
④
The FLS-1000 operates by generating steam under pressure and temperature. Therefore, its
sterilization process can be greatly affected by the atmospheric pressure at the location where
the FLS-1000 is installed. In order to ensure proper operation of the FLS-1000 please reset the
altitude level if it is moved to another location.
Table. Altitude level code and the corresponding altitude level and maximum
setting sterilization and lid interlock releasing temperatures.
Altitude level code
Altitude level
Maximum setting
sterilization temp.
Maximum lid interlock
releasing temp.
0
Below 800m
123℃
100℃
1
Above 800m
〜
below 1500m
122℃
97℃
2
Above 1500m
〜
below 2000m
121℃
95℃
Press and hold down the START key.
The buzzer will sound in approximately 4 seconds to indicate that the setting is finalized and
the display will return to the normal.
⑤
STOP
MEMORY
STERILIZE
LIQUID
START
FUNCTION
CHECK
TIMER
EXT.
SENSOR
COOLING
FAN
EXHAUST SPEED
FAST
TIME
TEMPERATURE
TIMER
READY
STERILIZE
LID
LID
WATER
DRAIN
COVER
CHANGE WATER
°C
SLOW
STERILIZE
IN PROGRESS
STOP
MEMORY
STERILIZE
LIQUID
START
FUNCTION
CHECK
TIMER
EXT.
SENSOR
COOLING
FAN
EXHAUST SPEED
FAST
TIME
TEMPERATURE
TIMER
READY
COMPLETED
POWER
FAILURE
STERILIZE
LID
LID
WATER
DRAIN
COVER
CHANGE WATER
°C
SLOW
STERILIZE
IN PROGRESS
Summary of Contents for FLS-1000
Page 9: ...ー 8 ー MEMO ...
Page 10: ...ー 9 ー 2 2 Parts Identifications FLS 1000 Main Unit ...
Page 51: ...ー 50 ー MEMO ...