6-5. Power Failure Alert Function
If a power failure occurs or the autoclave is turned off at the power switch during operation, the
POWER FAILURE indicator will turn on when the power is restored. If a power failure occurs
before the sterilization process has completed, the time display will show the error code “A09”
to indicate that the chamber is not heated when the power returns. Operation can be restarted
while the error code “A09” is displayed. If a power failure occurs after the sterilization process
has completed, the autoclave will operate the temperature and pressure falling process or the
temperature falling process when the power returns.
Check that the error code “A09” appears in the
time display.
Check the operation process of the sterilization
cycle when a power failure occurred.
The process in operation when a power failure
occurred flashes red on the work monitor.
Press the START key to restart operation.
Press the STOP key to stop operation.
※
The cycle process to be resumed is determined
based on the information recorded before
interruption, and the data on temperature and
pressure at the time of resumption of operation.
※
The POWER FAILURE indicator turns off when the
lid is opened next time.
※
When the operation is restarted by this function, the
COMPLETED indicator will not turn on after the
operation is completed.
<
When the error code “A09”appears in the display
>
①
②
③
STOP
MEMORY
STERILIZE
LIQUID
START
FUNCTION
CHECK
TIMER
EXT.
SENSOR
COOLING
FAN
EXHAUST SPEED
FAST
TIME
TEMPERATURE
TIMER
POWER
FAILURE
STERILIZE
LID
LID
WATER
DRAIN
COVER
CHANGE WATER
°C
SLOW
STERILIZE
IN PROGRESS
STOP
MEMORY
STERILIZE
LIQUID
START
FUNCTION
CHECK
TIMER
EXT.
SENSOR
COOLING
FAN
EXHAUST SPEED
FAST
TIME
TEMPERATURE
TIMER
POWER
FAILURE
STERILIZE
LID
LID
WATER
DRAIN
COVER
CHANGE WATER
°C
SLOW
STERILIZE
IN PROGRESS
※ 1
After resumption the set sterilization time will be achieved regardless of the time already elapsed in
the sterilization process when a power failure occurred.
Operation process at the time of
interruption
Temperature and pressure at the time of resumption of
operation
Process to be resumed
Sterilization
Temperature over the set sterilization temperature
Sterilization
※
1
Positive pressure and temperature lower than the set
sterilization temperature
Temperature and pressure rise
Negative pressure
Temperature rise
Temperature and pressure rise
Positive pressure
Temperature and pressure rise
Negative pressure
Temperature rise
Temperature rise
-
Temperature rise
Temperature and pressure fall
Positive pressure
Temperature and pressure fall
Negative pressure
Temperature fall
Temperature fall
-
Temperature fall
Table. Basis for determination of the sterilization cycle process to be resumed