9-2. Error Code Table
Error code
Possible cause
Remedy
E04
Level of sterilizing water becomes
insufficient during the operation.
Turn off the power switch and wait until the
chamber pressure becomes 0MPa to open the lid.
And then pouring water into the chamber.
E05
Temperature inside the chamber is much
higher than the set temperature during the
sterilization process.
※
It exceeds the set value by 5℃ or more.
Unplug the power plug of the FLS-1000 from
the outlet and contact your dealer or our local
representative.
E06
Pressure inside the chamber rises because
the shape of the article to be sterilized serves
to prevent air from escaping.
※
The pressure exceeds 175kPa.
Turn off the power switch and wait until the
chamber pressure becomes 0MPa to open the
lid. And then take some measures to allow air to
escape from the load or reduce the volume of the
load.
(Refer to “5-6. Placing Articles to be Sterilized”.)
E07
Chamber temperature sensor is disconnected
or broken.
Unplug the power plug of the FLS-1000 from
the outlet and contact your dealer or our local
representative.
E08
The temperature inside the chamber is much
lower than the set temperature during the
sterilization process.
※
It is more than 5℃ lower than the set
value.
Unplug the power plug of the FLS-1000 from
the outlet and contact your dealer or our local
representative.
E09
Chamber lid opens during the operation.
※
Lid open is detected by the sensor.
Close the chamber lid securely. If this happens
during the operation, unplug the power plug of
the FLS-1000 from the outlet and contact your
dealer or our local representative.
E10
Interlocking system of the chamber lid is not
activated.
Unplug the power plug of the FLS-1000 from
the outlet and contact your dealer or our local
representative.
E11
Temperature does not rise during
temperature rising and temperature and
pressure rising processes.
※
Temperature rises less than 2℃ within
the specified time limit. A temperature
equal to or below 60℃ in 120 minutes,
and above 60℃ in 40 minutes.
Unplug the power plug of the FLS-1000 from
the outlet and contact your dealer or our local
representative.
E12
Pressure sensor is disconnected or broken.
Unplug the power plug of the FLS-1000 from
the outlet and contact your dealer or our local
representative.
E13
External temperature sensor for articles to
be sterilized is disconnected or broken.
Unplug the power plug of the FLS-1000 from
the outlet and contact your dealer or our local
representative.
E15
Air vent valve is broken.
Unplug the power plug of the FLS-1000 from
the outlet and contact your dealer or our local
representative.
<