Operation
Diagnostics Menu
Thermo Electron Corporation
Model 42
i
Low Source Instruction Manual
3-71
Set Analog Outputs
The Set Analog Outputs screen contains three choices: Set to full-scale, set
to zero, or reset to normal. Full-scale sets the analog outputs to the full-scale
voltage, zero sets the analog outputs to 0 volts, and normal operation. The
example below shows the selected output state “ALL” is set to normal.
•
In the Main Menu, choose Diagnostics > Test Analog Outputs >
ALL,
Voltage Channel 1-6,
or
Current Channel 1-6
.
•
Use
and
to move the cursor up and down.
•
Press
to select a choice.
TEST ANALOG OUTPUTS:
>ALL
VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
VOLTAGE CHANNEL 5
VOLTAGE CHANNEL 6
ª
RANGE AVG DIAGS ALARM
SET ANALOG OUTPUTS:
SETTING: ALL
OUTPUT SET TO: NORMAL
©
SET TO FULL SCALE
ª
SET TO ZERO
§
RESET TO NORMAL
RANGE AVG DIAGS ALARM