Operation
Calibration Menu
Thermo Electron Corporation
Model 42
i
Low Source Instruction Manual
3-27
Next Time
The Next Time screen is used to view and set the next zero/span check date
and time. Once the initial zero/span check is performed, the date and time
of the next zero/span check is calculated and displayed.
•
In the Main Menu, choose Calibration > Zero/Span Check >
Next
Date/Time
.
•
Use
,
,
and
to move and change the value of
the date and time.
•
Press
to accept a change.
Period Hours
The Zero/Span Period Hours screen defines the period or interval between
zero/span checks. Periods between 0 and 999 hours are acceptable. To turn
the zero/span check off, set the period to 0.
•
In the Main Menu, choose Calibration > Zero/Span Check >
Period
Hours
.
•
Use
and
to increment and decrement the numeric value.
•
Press
to save the period.
NEXT DATE AND TIME:
NEXT DATE AND TIME:
01 JAN 2005 12:00
02 JAN 2005 12:00 ?
PRESS
TO EDIT
SETTING: DAYS
¨
SET MONTHS
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
RANGE AVG DIAGS ALARM
ZERO/SPAN PERIOD:
CURRENTLY: 024 HRS
SET TO: 02
5
HRS ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
Summary of Contents for 42i
Page 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Page 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 314: ......