Operation
Calibration Factors Menu
3-22
Model 42
i
Low Source Instruction Manual
Thermo Electron Corporation
Calibration Pressure
The Calibration Pressure screen reports the reactor pressure at which current
calibration was performed. The screen also allows this value to be edited by
the operator. This value should not be changed unless a complete manual
calibration is being performed. The calibration pressure is set equal to the
reactor pressure at the time of calibration.
Note
Turning the pressure compensation on and off can produce significant
artificial jumps in the apparent sample concentration. If the pressure
compensation feature is to be used, the instrument must be calibrated with
the pressure compensation feature on. For more information about
calibration, see Chapter 4, “Calibration”.
•
In the Main Menu, choose Calibration Factors
> Calibration Pressure
.
•
Use
and
to increment or decrement the calibration
pressure value.
•
Press
to save the new calibration pressure.
Reset User Calibration Default
The Reset User Calibration Default screen allows the user to reset the
calibration configuration values to factory defaults.
•
In the Main Menu, choose Service > Calibration Factors >
Reset User
Cal Defaults
.
•
Press
to warn user and enable restore with
.
•
Use
to reset the calibration pressure reference value when pressed
after .
CALIBRATION PRESSURE:
PRESSURE: 151.6
SET CAL PRES TO: 150.0
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
Summary of Contents for 42i
Page 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Page 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 314: ......