C-Link Protocol Commands
Alarms
Thermo Electron Corporation
Model 42
i
Low Source Instruction Manual
B-15
Send:
pressure alarm min
Receive:
pressure alarm min 150 mmHg
set alarm pressure min
value
set alarm pressure max
value
These commands set the pressure alarm minimum and maximum values to
value,
where
value
is a floating-point number representing pressure alarm
limits in millimeters of mercury. The example below sets the pressure alarm
maximum value to 290 mmHg.
Send:
set alarm pressure max 290
Receive:
set alarm pressure max 290 ok
alarm trig conc no
alarm trig conc no2
alarm trig conc nox
These commands report the NO, NO
2
, and NO
x
concentrations alarm
trigger action for minimum alarm, current setting, to either floor or ceiling.
The example below shows the NO concentration minimum alarm trigger to
ceiling, according to Table B-3.
Send:
alarm trig conc no
Receive:
alarm trig conc no 1
set alarm trig conc no
value
set alarm trig conc no2
value
set alarm trig conc nox
value
These commands set the NO, NO
2
, and NO
x
concentrations alarm
minimum
value,
where
value
is set to either floor or ceiling, according to
Table B-3. The example below sets the NO concentration minimum alarm
trigger to ceiling.
Send:
set alarm trig conc no 1
Receive:
set alarm trig conc no 1 ok
Table B-3.
Alarm Trigger Values
Value
Alarm Trigger
0
Floor
1
Ceiling
Summary of Contents for 42i
Page 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Page 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 314: ......