C-Link Protocol Commands
I/O Configuration
B-40
Model 42
i
Low Source Instruction Manual
Thermo Electron Corporation
Send:
set analog iout range 4 1
Receive:
set analog iout range 4 1 ok
analog vin
channel
This command retrieves the analog voltage input
channel
data, both the
calculated value and the actual voltage. In the example below, the
“calculated” value of channel 1 is 75.325 degrees F, volts are 2.796. This
command responds with “feature not enabled” if the I/O expansion board is
not detected.
Send:
analog vin 1
Receive:
analog vin 1 75.325 2.796
analog vout range
channel
This command reports the analog voltage output
channel
range, where
channel
is between 1 and 6 inclusive, according to Table B-11.
Send:
analog vout range 2
Receive:
analog vout range 2 3
set analog vout range
channel range
This command sets analog voltage output
channel
to the range, where
channel
is between 1 and 6 inclusive, and
range
is set according to Table
B-11. The example below sets channel 2 to the 0-10 V range.
Send:
set analog vout range 2 3
Receive:
set analog vout range 2 3 ok
Table B-10.
Analog Current Output Range Values
Range
Output Range
1
0-20 mA
2
4-20 mA
0 [cannot be set to this, but may report]
Undefined
Table B-11.
Analog Voltage Output Range Values
Range
Output Range
1
0-1 V
2
0-100 mV
3
0-10 V
4
0-5 V
0 [cannot be set to this, but may report]
Undefined
Summary of Contents for 42i
Page 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Page 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 314: ......