![Thermo 42i Instruction Manual Download Page 283](http://html.mh-extra.com/html/thermo/42i/42i_instruction-manual_1104217283.webp)
C-Link Protocol Commands
Communications Configuration
Thermo Electron Corporation
Model 42
i
Low Source Instruction Manual
B-35
save
set save params
This command stores all current parameters in FLASH memory. It is
important that each time instrument parameters are changed, that this
command be sent. If changes are not saved, they will be lost in the event of
a power failure. The example below saves the parameters to FLASH
memory.
Send:
set save params
Receive:
set save params ok
time
This command reports the current time (24-hour time). The example
below reports that the internal time is 2:15:30 pm.
Send:
time
Receive:
time 14:15:30
set time
hh:mm:ss
hh
= hours
mm
= minutes
ss
= seconds
This command sets the internal clock (24-hour time). The example below
sets the internal time to 2:15 pm.
Note
if seconds are omitted, the seconds default to 00.
Send:
set time 14:15
Receive:
set time 14:15 ok
Communications
Configuration
addr dns
This command reports the TCP/IP address for the domain name server.
Send:
addr dns
Receive:
addr dns 192.168.1.1
set addr dns
address
This command sets the domain name server
address
, where
address
consists
of four numbers ranging from 0-255 inclusive, separated by “.”.
Send:
set addr dns 192.168.1.1
Receive:
set addr dns 192.168.1.1 ok
Summary of Contents for 42i
Page 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Page 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 314: ......