Operation
Instrument Controls Menu
3-42
Model 42
i
Low Source Instruction Manual
Thermo Electron Corporation
Communication Settings
The communication Settings menu is used with communications control
and configuration.
•
In the Main Menu, choose Instrument Controls >
Communication
Settings
.
Baud Rate
The Baud Rate screen is used to set the RS-232/RS-485 interface baud rate.
Baud rates of 1200, 2400, 4800, and 9600, 19200, 38400, 57600, and
115200 are available.
•
In the Main Menu, choose Instrument Controls > Communication
Settings >
Baud Rate
.
•
Use
and
to scroll through a list of choices.
•
Press
to save the new baud rate.
Instrument ID
The Instrument ID screen allows the operator to edit the instrument ID.
The ID is used to identify the instrument when using the C-Link or
MODBUS protocols to control the instrument or collect data. It may be
necessary to edit the ID number if two or more of the same instrument are
connected to one computer over the serial port. Valid Instrument ID
numbers are from 0 to 127. The Model 42
i
Low Source has a default
COMMUNICATION SETTINGS:
>BAUD RATE
INSTRUMENT ID
COMMUNICATION PROTOCOL
STREAMING DATA CONFIG
RS-232/RS-485 SELECTION
TCP/IP SETTINGS
RANGE AVG DIAGS ALARM
BAUD RATE:
CURRENTLY: 9600
SET TO: 19200 ?
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
Summary of Contents for 42i
Page 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Page 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 314: ......