Servicing
PMT Cooler and Reaction Chamber Assembly Replacement
7-14
Model 42
i
Low Source Instruction Manual
Thermo Electron Corporation
1. Refer to “Removing the Measurement Bench and Lowering the
Partition Panel” in this chapter to lower the partition panel, then
proceed to the next step below.
2. Disconnect the reaction chamber connector from the temperature
control board.
3. Snap off the temperature control board from the board mounts.
4. Remove the four screws securing the cooler shroud and remove the
shroud.
Figure 7-8.
PMT Cooler and Reaction Chamber
5. Disconnect plumbing connections from the reaction chamber.
6. Disconnect the PMT cooler cables from the PMT high voltage power
supply, the input board, and the measurement interface board. Remove
all tie-wraps securing the cables.
Captive Screws (4)
Cooler Shroud
Temperature Control Board
Reaction Chamber
BNC Connector to Input Board
To PMT HVPS
To BENCH on
To COOLER on Measurement
Interface Board
Temp. Control Board
(Two Not Shown)