Servicing
DC Power Supply Replacement
Thermo Electron Corporation
Model 42
i
Low Source Instruction Manual
7-29
4. At the GAIN 100 screen, use
to increment the D/A
counts until the FREQ value matches or is slightly above (within 50
counts) the value noted in the previous step.
5. Press
to store the value.
The screen flashes
Calculating - Please Wait!
and
Done - Values
Saved!
messages.
DC Power Supply
Replacement
Use the following procedure to replace the DC power supply (
Equipment Required:
DC power supply
Philips screwdriver
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Disconnect all the power supply electrical connections. Note connector
locations to facilitate re-connection.
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.