C-Link Protocol Commands
I/O Configuration
Thermo Electron Corporation
Model 42
i
Low Source Instruction Manual
B-43
Send:
list var aout
Receive:
list var aout
index variable
0 none
1 no
2 no2
3 nox
18 intt
19 rctt
20 pmtt
21 convt
25 pres
27 pmtv
28 ain1
29 ain2
30 ain3
31 ain4
32 ain5
33 ain6
34 ain7
35 ain8
relay stat
This command reports the current relay logic normally “open” or normally
“closed,” if all the relays are set to same state, that is all open or all closed.
The example below shows that the status when all the relays logic is set to
normally “open”.
Send:
relay stat
Receive:
relay stat open
Note
if individual relays have been assigned different logic then the response
would be a 4-digit hexadecimal string with the least significant byte (LSB)
being relay no 1.
For example:
Receive:
relay stat 0x0001 (indicates relay no 1 is set to normally
open logic, all others are normally closed)
Receive:
relay stat 0x0005 (indicates relay no 1 and 3 are set to be
normally open logic, all others are normally closed)
set relay open
set relay open
value
set relay closed
set relay closed
value
These commands set the relay logic to normally open or closed for relay
number
value
, where
value
is the relay between 1 and 16. The example
below sets the relay no 1 logic to normally open.
Summary of Contents for 42i
Page 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Page 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 314: ......