C-Link Protocol Commands
Communications Configuration
B-36
Model 42
i
Low Source Instruction Manual
Thermo Electron Corporation
addr gw
This command reports the default TCP/IP gateway address.
Send:
addr gw
Receive:
addr gw 192.168.1.1
set addr gw
address
This command sets the default gateway
address
, where
address
consists of
four numbers ranging from 0-255 inclusive, separated by “.”.
Send:
set addr gw 192.168.1.1
Receive:
set addr gw 192.168.1.1 ok
addr ip
This command reports the IP address of the analyzer.
Send:
addr ip
Receive:
addr ip 192.168.1.15
set addr ip
address
This command sets the analyzer’s IP
address
, where
address
consists of four
numbers ranging from 0-255 inclusive, separated by “.”.
Send:
set addr ip 192.168.1.15
Receive:
set addr ip 192.168.1.15 ok
addr nm
This command reports the TCP/IP netmask address.
Send:
addr nm
Receive:
addr nm 255.255.255.0
set addr nm
address
This command sets the netmask
address
, where
address
consists of four
numbers ranging from 0-255 inclusive, separated by “.”.
Send:
set addr nm 255.255.255.0
Receive:
set addr nm 255.255.255.0 ok
baud
This command reports the current baud rate for the serial port
(RS232/RS485). The example below reports that the current baud rate is
9600.
Send:
baud
Receive:
baud 9600
Summary of Contents for 42i
Page 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Page 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 314: ......