Servicing
Photomultiplier Tube Replacement
Thermo Electron Corporation
Model 42
i
Low Source Instruction Manual
7-15
7. Loosen four captive screws holding cooler to floor plate and remove the
cooler assembly with the reaction chamber.
Note
If only the cooler is being replaced, remove the PMT and reaction
chamber from the old cooler and install them on the new cooler.
Fasten knurled fittings on reaction chamber finger tight.
Make sure that the heat shrink covered tubing between the reaction
chamber and the converter is light tight at the connections.
8. Install new cooler by following previous steps in reverse.
9. Re-install the measurement bench. Refer to “Removing the
Measurement Bench and Lowering the Partition Panel” in this chapter.
Photomultiplier Tube
Replacement
Use the following procedure to replace the PMT tube.
Equipment Required:
Photomultiplier tube and PMT base
Nut driver, 5/16-inch
Flat blade screwdriver
Philips screwdriver, small
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Disconnect the high voltage cable from the PMT power supply and
unplug the BNC cable from the Input Board.
3. Remove six external screws holding PMT cover plate and the four
screws holding the PMT shroud to the panel and remove the PMT
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
Summary of Contents for 42i
Page 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Page 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 314: ......