Servicing
Analog Output Testing
Thermo Electron Corporation
Model 42
i
Low Source Instruction Manual
7-31
Figure 7-17.
Rear Panel Analog Voltage Output Pins
2. From the Main Menu,
press
to scroll to Diagnostics, > press
>
to scroll to Test Analog Outputs, and press
.
The Test Analog Outputs screen appears.
3. Press
to scroll to the channel (Voltage Channel 1-6)
corresponding to the rear panel terminal pin where the meter is
connected, and press
.
Table 7-3.
Analog Output Channels and Rear Panel Pin Connections
Channel
Pin
1
14
2
33
3
15
4
34
5
17
6
36
Ground
16, 18, 19, 35, 37