![Thermo 42i Instruction Manual Download Page 297](http://html.mh-extra.com/html/thermo/42i/42i_instruction-manual_1104217297.webp)
Thermo Electron Corporation
42
i
Low Source Instruction Manual
C-1
Appendix C
MODBUS Protocol
This appendix provides a description of the MODBUS Protocol Interface
and is supported both over RS-232/485 (RTU protocol) as well as TCP/IP
over Ethernet.
The MODBUS Commands that are implemented are explained in detail in
this document. The MODBUS protocol support for the
i
Series enables the
user to perform the functions of reading the various concentrations and
other analog values or variables, read the status of the digital outputs of the
analyzer, and to trigger or simulate the activation of a digital input to the
instrument. This is achieved by using the supported MODBUS commands
listed below.
For details of the Model 42
i
Low Source MODBUS Protocol specification,
see the following topics:
•
“Serial Communication Parameters”
parameters used to support MODBUS RTU protocol.
•
“TCP Communication Parameters”
parameters used for TCP connection.
•
“Application Data Unit Definition”
used over serial and TCP/IP.
•
describes the various function codes
supported by the instrument.
•
commands supported in Table C-1 through Table C-3.
Additional information on the MODBUS protocol can be obtained at
. References are from MODBUS Application
Protocol Specification V1.1a MODBUS-IDA June 4, 2004.
Summary of Contents for 42i
Page 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Page 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 314: ......