C-Link Protocol Commands
Measurements
Thermo Electron Corporation
Model 42
i
Low Source Instruction Manual
B-11
pmt voltage
This command reports the current PMT voltage. The example below
reports that the current PMT voltage is -750 volts.
Send:
pmt voltage
Receive:
pmt voltage -750
pres
This command reports the current reaction chamber pressure. The example
below shows that the actual reaction chamber pressure is 232.5 mmHg.
Send:
pres
Receive:
pres 232.5 mm Hg
react temp
This command reports the current reaction chamber temperature. The
example below reports that the current reaction chamber temperature is
49.0 °C.
Send:
react temp
Receive:
react temp 49.0 deg C
flags
This reports 8 hexadecimal digits (or flags) that represent status of the
ozonator, PMT, pressure and temperature compensation status, gas units,
gas mode, and alarms. To decode the flags, each hexadecimal digit is
converted to binary as shown in the Figure B-1. It is the binary digits that
define the status of each parameter. In the example below, the instrument is
reporting that the ozonator and PMT are both on, that the instrument is in
the span gas mode, that at least one of the alarm temperature is activated,
and that the NO
2
high concentration alarm is activated.
Send:
flags
Receive:
flags C0028008
Summary of Contents for 42i
Page 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Page 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 314: ......