Calibration
Calibration in Dual Range and Auto Range Mode
Thermo Electron Corporation
Model 42
i
Low Source Instruction Manual
4-13
5. Press
to set the NO background to zero.
6. Press
to return to the Run screen.
7. Repeat Steps 3 through 6, selecting
Calibrate NOx Background
to set
the NO
x
background to zero.
Calibrate Low NO
Use the following procedure to calibrate NO channel to the NO calibration
gas.
1. Disconnect the source of zero air from the SAMPLE
bulkhead. In its
place, connect a source of NO calibration gas of about 80% of the low
NO full-scale range.
2. Allow the analyzer to sample the low NO calibration gas until the NO,
NO
2
, and NO
x
readings stabilize.
3. When the responses are stabile, from the Main Menu choose
Calibration >
Calibrate Lo RNG NO
.
4. The Lo NO field displays the current NO concentration.
The Lo NO Span Conc field is where you enter the low NO calibration
gas concentration.
Use
to move the cursor left and right and use
to increment and decrement the numeric character at the cursor.
5. Press
to calculate and save the new low NO coefficient based on
the entered span concentration.
Calibrate Low NO
x
Use the following procedure to calibrate the NO
x
channel to the NO
x
calibration gas.
1. Press
to return to the Calibration menu and choose
Calibrate
Lo RNG NO
x
.
Summary of Contents for 42i
Page 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Page 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 314: ......