Operation
Instrument Controls Menu
Thermo Electron Corporation
Model 42
i
Low Source Instruction Manual
3-63
Pressure Compensation
Pressure compensation provides compensation for any changes to the
instrument's output signal due to reaction chamber pressure variations. The
effects of reaction chamber pressure changes on the analyzer's subsystems
and output have been empirically determined. This empirical data is used to
compensate for any change in reaction chamber pressure. This
compensation can be used even though the Model 42
i
Low Source does not
require pressure compensation for EPA equivalency.
When pressure compensation is on, the first line of the display represents
the current pressure in the reaction chamber. When pressure compensation
is off, the first line of the display shows the factory standard pressure of 150
mmHg.
•
In the Main Menu, choose Instrument Controls >
Pressure
Compensation
.
•
Press
to toggle and set the pressure compensation on or off.
TEMPERATURE COMPENSATION:
COMP TEMP: 30.0
o
C
CURRENTLY: OFF
SET TO: ON ?
TOGGLE VALUE
RANGE AVG DIAGS ALARM
PRESSURE COMPENSATION:
COMP PRES: 150.0 mmHg
CURRENTLY: OFF
SET TO: ON ?
TOGGLE VALUE
RANGE AVG DIAGS ALARM
Summary of Contents for 42i
Page 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Page 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 314: ......