![Thermo 42i Instruction Manual Download Page 262](http://html.mh-extra.com/html/thermo/42i/42i_instruction-manual_1104217262.webp)
C-Link Protocol Commands
Alarms
B-14
Model 42
i
Low Source Instruction Manual
Thermo Electron Corporation
alarm cooler temp min
alarm cooler temp max
These commands report the cooler temperature alarm minimum and
maximum value current settings. The example below reports that the cooler
temperature alarm minimum value is -10.0 °C.
Send:
alarm cooler temp min
Receive:
alarm cooler temp min -10.0 deg C
set alarm cooler temp min
value
set alarm cooler temp max
value
These commands set the cooler temperature alarm minimum and
maximum values to
value
, where
value
is a floating-point number
representing cooler temperature alarm limits in degrees C. The example
below sets the cooler temperature alarm maximum value to -2.0 °C.
Send:
set alarm cooler temp max -2
Receive:
set alarm cooler temp max -2 ok
alarm internal temp min
alarm internal temp max
These commands report the internal temperature alarm minimum and
maximum value current settings. The example below reports that the
internal temperature alarm minimum value is 15.0 °C.
Send:
internal temp alarm min
Receive:
internal temp alarm min 15.0 deg C
set internal temp alarm min
value
set internal temp alarm max
value
These commands set the internal temperature alarm minimum and
maximum values to
value
, where
value
is a floating-point number
representing internal temperature alarm limits in degrees C. The example
below sets the internal temperature alarm maximum value to 35.0 °C.
Send:
set internal temp alarm max 35
Receive:
set internal temp alarm max 35 ok
alarm pressure min
alarm pressure max
These commands report the pressure alarm minimum and maximum value
current settings. The example below reports that the pressure alarm
minimum value is 150 mmHg.
Summary of Contents for 42i
Page 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Page 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 314: ......