Operation
Alarms Menu
3-78
Model 42
i
Low Source Instruction Manual
Thermo Electron Corporation
Pressure
The Pressure screen displays the current reaction chamber pressure reading
and sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 100 to 200 mmHg. If the pressure reading goes beyond either
the minimum or maximum limit, an alarm is activated. The word
“ALARM” appears in the Run screen and in the Main Menu.
•
In the Main Menu, choose Alarms >
Pressure
.
Min and Max Pressure Limits
The Minimum Pressure alarm limit screen is used to change the minimum
pressure alarm limit. The minimum and maximum pressure screens
function the same way.
•
In the Main Menu, choose Alarms > Pressure >
Min
or
Max
.
•
Use
and
to increment or decrement the numeric value.
•
Press
to save set to value as actual value.
PRESSURE:
ACTUAL 151.6 mmHg
>MIN 100.0 mmHg
MAX 200.0 mmHg
RANGE AVG DIAGS ALARM
PRESSURE:
ACTUAL MIN: 100.0 mmHg
SET MIN TO: 110.0 mmHg?
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM