Operation
Alarms Menu
Thermo Electron Corporation
Model 42
i
Low Source Instruction Manual
3-81
Min and Max NO, NO
2
, and NO
x
Concentration Limits
The Minimum NO Concentration alarm limit screen is used to change the
minimum NO concentration alarm limit. The minimum and maximum
NO, NO
2
, and NO
x
concentration alarm limit screens function the same
way.
•
In the Main Menu, choose Alarms > Select Concentration >
Min
or
Max
.
•
Use
and
to increment or decrement the numeric value.
•
Press
to save set to value as actual value.
Min Trigger
The Minimum Trigger screen allows the user to view and set the NO, NO
2
,
and NO
x
concentration low alarm trigger type to either floor or ceiling. The
minimum alarm may be programmed as a floor trigger (alarm is triggered
when the concentration falls below the minimum value) or a ceiling trigger
(alarm is triggered when the concentration goes above the minimum value).
•
In the Main Menu, choose Alarms > Select Concentration >
MinTrigger
.
•
Press
to toggle and set to floor or ceiling.
NO CONCENTRATION:
ACTUAL 62.7
>MIN 0
MAX 500000
MIN TRIGGER CEILING
RANGE AVG DIAGS ALARM
NO CONCENTRATION:
ACTUAL MIN: 0
SET MIN TO: 0000000
1
?
§¨
MOVE CURSOR
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
Summary of Contents for 42i
Page 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Page 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Page 314: ......