Optional Equipment
Permeation Dryer Assembly
9-2
Model 42
i
Low Source Instruction Manual
Thermo Electron Corporation
Figure 9-1.
Flow Diagram, Zero/Span Option
Permeation Dryer
Assembly
The permeation dryer minimizes routine maintenance procedures by
providing a continuous stream of dry air to the ozonator (using the selective
water permeation characteristics of the dryer). With the permeation dryer
option, it is not necessary to constantly replenish the ozonator air-drying
column as in the standard instrument.
If this option is installed, refer to the “Operation” chapter. for more
information.
Teflon Particulate
Filter
A 5-10 micron pore size, two-inch diameter Teflon® element is available for
the Model 42
i
Low Source. This filter should be installed just prior to the
SAMPLE
bulkhead. When using a filter, all calibrations and span checks
must be performed through the filter.
Ozone Particulate
Filter
The ozone particulate filter minimizes the potential for contamination of
the ozonator by trapping any particulate matter before passing through the
ozonator.