VI-7
Troubleshooting
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
The .Thermo .Scientific .Orion .2111LL .sodium .monitor .allows .the .user .to .
reset .the .monitor .through .the .setup .mode .or .by .a .hard .reset . . .Resetting .the .
monitor .will .restore .all .settings .in .the .setup .menu .to .factory .default .values . .
(See .ISE .Default .Values)
Warning: .Resetting .the .monitor .will .erase .all .data .logs .including .
calibration, .and .password .protection .settings .for .setup .and .calibration . . . .
The .instrument .will .reset .setup .parameters .to .factory .default .settings .
User Action
Marquee
Scrolling
Main Display
top line
bottom line
Notes
Press
.
to enter setup mode.
Last menu used
in setup mode
SEL
SCrn
SETUP appears
as mode
Press
.
./ .
until RSET
appears in the Marquee.
RSET
(Flashing)
SEL
SCrn
Press
.
.
PRESS TEST
VIEW ENTER TO
RESET
rSEt
?
Press in the following sequence:
1.
.
2.
.
.
3.
.
.
System resets to default values
and returns to the measurement
mode. Actual values for
concentration and temperature
will be displayed if the electrodes
are operational.
PRESS TEST
VIEW ENTER TO
RESET
rSEt
?
Resetting the
Monitor