III-9
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default
User Action
Marquee
Scrolling
Main Display
top line
bottom line
Notes
DKA (cont’d)
Press
to accept value and advance
to next screen to set concentration for
Standard 2
9 .0 ppm
SET STD 2 CONCENTRATION
CONC
# # #
(Flashing)
To set value for Standard 2 concentration
(ppm or ppb):
Press
to move decimal and ppm/ppb
Press
/
to set values of digits
Press
to move to set each digit
SET STD 2 CONCENTRATION
CONC
#
#. #
(Flashing digit will change)
Press
to accept values and advance
to next screen to set volume for Standard 2
0 .0 mL
SET STD 2 VOLUME IN mL
Std 2
#. # #
(Flashing)
To set value for Standard 2 volume (mL):
Press
/
to set values of digits
Press
to move to set each digit
SET STD 2 VOLUME IN mL
Std 2
#
#. #
(Flashing digit will
change)
Press
to return user to main setup
menu for DKA submenu
DKA
(Flashing)
SEL
SCrn
Press
to scroll to next menu