III-
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default User Action
Marquee
Scrolling
Main Display
top line
bottom line
Notes
mAMP
mAMP
(flashing)
SEL
SCrn
Press
to set analog output parameters
4-0
Select 0-20 or 4-20
4-20
(Flashing)
Press
/
to toggle between 4-20 or
0-20 output values
Select 0-20 or 4-20
4-20 or 0-20
(Flashing)
Press
to set analog output and advance to
next screen for sensing output (SOUt) for sodium
LOG
Press
/
to toggle
between LOg and LIn output
SELECT LOG OR LINEAR
OUTPUT
SOUt
LOg or LIn
(Flashing)
The default
scale values
the same for
both LOg and
LIn
Press
to set SOUt low (LO) limit
.0 ppb
mA S ENSOR OUTPUT LOW
VALUE
LO
#.# ppb
(Flashing)
Press
to move decimal and ppm / ppb
Press
/
to set digit values
mA SENSOR OUTPUT LOW
VALUE
LO
#
# #. #
(Flashing digit will change)
Press
to move to set each digit
Press
/
to set values of remaining
digits for SOUt low value
mA SENSOR OUTPUT LOW
VALUE
LO
# # #.
#
(Flashing digit will change)
Press
to accept value and advance to next
screen to set SOUt high (HI) limit
00 ppb
mA SENSOR OUTPUT HIGH
VALUE
HI
#.# ppb
(Flashing)
Press
to move decimal and ppm / ppb
Press
/
to set digit values
mA SENSOR OUTPUT HIGH
VALUE
HI
#
# #. #
(Flashing digit will change)
Press
to move to set each digit
Press
/
to set values of remaining
digits for SOUt high value
mA SENSOR OUTPUT HIGH
VALUE
HI
###.
#
(Flashing digits will
change)