IV-9
Calibration
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Stage
User Action
Marquee
Main Display
top line
bottom line
Valve position
7
Wait 0 - 5 minutes
AWAITING DATA
STABILITY
Actual readings in ppb
& mV flashing
System continuously evaluates
data by applying stability criteria
until a “stable/pass” condition is
met. If system stabilizes follow
stage 8
.
If system stability is not reached
a timeout occurs. Error E041 is
issued. Jump to step
Error E04
.
8
No action necessary
System calculates new Eo and
slope and checks for errors. If
no errors are found follow
stage 9
.
If there is a slope/E
0
or other
cal data related error, jump to
appropriate error code.
9
Press
to accept the new
slope and Eo values
or
Press
to abort calibration.
The values will not be stored. The
last saved values for calibration
prior to the aborted values will be
used in measurement mode.
NEW CALIBRATION
SLOPE AND E
0
PRESS ENTER TO
ACCEPT OR PRESS EXIT
TO ABORT WITHOUT
SAVING.
Calculated Slope and E
0
System presents the new slope
and E
0
values.
If
is pressed, continue to
follow
stage 30
.
If
is pressed, refer to step
Exit
or
Exit 4
, depending on
the system software revision.
Some system software revisions
will not display the following
sections that are highlighted in
gray
NEW CALIBRATION
SLOPE AND E0 SAVED.
PRESS ENTER
CAL StOr
Calibration values for Slope and
Eo are stored.
Successful calibration
completed.
Press