II-6
Instrument Preparation
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Wiring the Instrument
Warning:
Be .sure .to .read .and .observe .the .following .requirements! . .If .you .
install .the .wrong .fuse .for .your .system, .you .could .damage .the .instrument . . .
Make .sure .you .select .the .correct .fuse .rating .and .discard .the .additional .fuses .
supplied .in .the .fuse .kit . . .
s
• . Options .kit- .fuses, .cable .glands, .conduit .fitting, .and .green .screw .
terminal .
• . Phillips .head .screwdriver .
• . 2 .mm .Blade .flat-head .screwdriver .
1 . . Open .the .faceplate: .Loosen .the .four .screws .using .a .Phillips .head .
screwdriver . . .Electronics .faceplate .will .open .via .the .hinge .pin .
connection .
2 . . Remove .one .or .two .of .the .two .unused .cable .glands .as .required .for .
wiring .power .cable .or .auxiliary .connections . . .Power .cable .optional .hole .
locations .in .
Figure II-4
.
3 . . Select .and .install .appropriate .size .cable .gland .or .conduit .fitting .as .
required .
4 . . Feed .power .cable .through .conduit .or .cable .glands .as .appropriate .
Figure II-4
Electronics Enclosure
with Cable Glands
4 Compression
Screws
Hinge Pin
Power Cable
Hole Locations
Cable Glands
Electrode Cables
Required Tools