VI-4
Troubleshooting
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Error/Event .codes .will .be .in .the .format .“E###” . . .Some .of .these .are .errors, .
some .are .warnings, .and .some .are .purely .informational . . .The .first .digits .
identify .the .type .of .error .or .event .as .identified .in .the .table .below .
Error/Event Codes
Displayed
Event Code
Description
Cause
Troubleshooting
System Status Codes
A7
ALARM1 CH1 ON
Alarm 1 Channel 1 engaged (closed)
A7
ALARM2 CH1 ON
Alarm 2 Channel 1 engaged (closed)
A73
ALARM3 CH1 ON
Alarm 3 Channel 1 engaged (open)
A74
AIR PUMP ON
Air pump engaged
A8
ALARM1 CH1 OFF
Alarm 1 Channel 1 disengaged (open)
A8
ALARM2 CH1 OFF
Alarm 2 Channel 1 disengaged (open)
A83
ALARM3 CH1 OFF
Alarm 3 Channel 1 disengaged (closed)
A84
AIR PUMP OFF
Air pump disengaged
C500
CAL MODE
Calibration mode entered
C509
CAL ISE DKA
DKA calibration complete
C50B
CAL ISE OFFLINE
Offline calibration complete
R400
MEAS MODE
Measurement mode entered
S600
SETUP MODE
Setup mode entered
E08
POWER FAIL
Brown out or power failure and system has
stopped processing.
Verify custom settings
E033
KEYPAD FAILURE
User did not press the keys during a
diagnostic mode keypad test, or keypad
broken
Repeat diagnostic mode keypad test
E034
CAL PASS INCORRECT
User entered incorrect calibration password
Verify password
E035
SYSTEM PASS INCORRECT
User entered incorrect system password
Verify password
E036
USER VALUE INCORRECT
User entered a value that is out of range
Verify value and re-enter
E037
POWER RETURN
System started processing after brown out
or power failure
Verify custom settings
E04
CALIBRATION TIMEOUT
The electrode has not reach stability within
the specified time
The user can press enter to force stability. The system
code may be due to out of range measurements or an
invalid previous calibration.
E04
NEGATIVE SLOPE
(INVALID SLOPE)
Invalid slope obtained during calibration
Check electrodes, electrode cables, and connectors for
defects or shorts. Replace if necessary. Recalibrate in the
correct sequence using standard 1 then standard 2.