III-4
Instrument Operation
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Default
User Action
Marquee
Scrolling
Main Display
top line
bottom line
Notes
TEST (cont’d)
To Select: SOUt
4-0
Press
/
to select 4-20
SELECT mA OR RELAY
tESt
4-20
(Flashing)
Press
to test 4-20 outputs
SELECT TEMP OR SENSOR OUTPUT tOUt or SOUt
(Flashing)
Press
/
to toggle between
SOUt and tOUt. Select SOUt
SELECT TEMP OR SENSOR OUTPUT SOUt
(Flashing)
Press
for SOUt; low value (LO)
Actual
LO Value
Displayed
mA SENSOR OUTPUT LOW VALUE
LO
#. # C
4 .0 mA are
sourced
at output
terminal 9
Press
for SOUt; high value (HI)
Actual
HI Value
Displayed
mA SENSOR OUTPUT HIGH VALUE
HI
#. # C
0 .0 mA are
sourced
at output
terminal 9
Press
to return user to main setup
menu for TEST submenu
TEST
(Flashing)
SEL
SCrn
Press
to test mAmp outputs (4-20),
relays (rLY), and the air pump (AIr)
To Select: rLY
RLY
Press
/
to select rLY
SELECT mA OR RELAY
tESt
rLY
(Flashing)
Press
to test rLY outputs